Aufsatz in einer Fachzeitschrift

Critical issues in conducting constant strain rate nanoindentation tests at higher strain rates



Details zur Publikation
Autor(inn)en:
Merle, B.; Higgins, W.; Pharr, G.
Verlag:
CAMBRIDGE UNIV PRESS

Publikationsjahr:
2019
Zeitschrift:
Journal of Materials Research
Seitenbereich:
3495-3503
Jahrgang/Band :
34
Heftnummer:
20
Erste Seite:
3495
Letzte Seite:
3503
Seitenumfang:
9
ISSN:
0884-2914
DOI-Link der Erstveröffentlichung:


Zusammenfassung, Abstract
Constant strain rate nanoindentation is a popular technique for probing the local mechanical properties of materials but is usually restricted to strain rates 0.1 pound s -1. Faster indentation potentially results in an overestimation of the hardness because of the plasticity error associated with the continuous stiffness measurement (CSM) method. This can have significant consequences in some applications, such as the measurement of strain rate sensitivity. The experimental strain rate range can be extended by increasing the harmonic frequency of the CSM oscillation. However, with commercial instruments, this is achievable only by identifying higher CSM frequencies at which the testing system is dynamically well behaved. Using these principles, a commercial system operated at the unusually high harmonic frequency of 1570 Hz was successfully used to characterize of the strain rate sensitivity of a Zn22Al superplastic alloy at strain rates up to 1 s(-1), i.e., an order of magnitude higher than with standard methods.


Schlagwörter
continuous stiffness measurement, nanoindentation, strain rate


Autor(inn)en / Herausgeber(innen)

Zuletzt aktualisiert 2022-20-12 um 13:09